14.03.2013 Views

Capnocheck® Capnometer - Smiths Medical

Capnocheck® Capnometer - Smiths Medical

Capnocheck® Capnometer - Smiths Medical

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Capnocheck ®<br />

PATienT MoniTorinG<br />

<strong>Capnometer</strong><br />

The BCi ® Capnocheck ®<br />

Capno meter is a small portable<br />

quantitative capnometer<br />

that monitors carbon dioxide<br />

concentrations and respiratory<br />

rate in one easy to use device<br />

that can be used for<br />

• Intubation verification<br />

• Return of spontaneous<br />

circulation during CPR<br />

• Routine airway management<br />

• Transport<br />

• Ventilator weaning<br />

Unlike colorimetric and other<br />

qualitative mechanical devices,<br />

the Capnocheck ® <strong>Capnometer</strong><br />

provides a fully quantitative value<br />

that is now recommended in the<br />

AHA 2005 guidelines. During<br />

resuscitation, the Capnocheck ®<br />

<strong>Capnometer</strong> can be an effective<br />

noninvasive indicator of cardiac<br />

output, and return of spontaneous<br />

circulation. The simple operation<br />

and accuracy of the Capnocheck ®<br />

<strong>Capnometer</strong> make it an invaluable<br />

tool in all areas of clinical practice.<br />

For more information, visit<br />

smiths-medical.com/capnocheck.


Alarm Silence Button (WW8110-001 only)<br />

Alarm Status Indicator<br />

(not illuminated)<br />

Breath Indicator Bar<br />

Battery Status Indicator<br />

Power Button<br />

BCi ®<br />

Capnocheck ®<br />

<strong>Capnometer</strong><br />

The ultra-compact Capnocheck ® <strong>Capnometer</strong> makes monitoring<br />

patients' ETCO 2 easy, efficient, and dependable.<br />

• Intuitive to use<br />

• Fully quantitative<br />

• Miniaturized mainstream device<br />

• Pocket size and lightweight (2.1 oz)<br />

• Battery powered – 2 AAA for<br />

continuous operation up to<br />

8 hours<br />

Ventilator Circuit VR1 Emergency and Transport Ventilator<br />

Bag Valve Mask<br />

examples of ventilation Sources<br />

Examples of Airway Connections<br />

Endotracheal Tube Mask<br />

Respiratory Rate Value<br />

Airway Adapter<br />

• No calibration required<br />

• No warm-up time<br />

• Two versions available –<br />

with or without alarms<br />

• Lightweight, small,<br />

easy storage<br />

ETCO 2 Value<br />

Laryngeal Mask


The Capnocheck ®<br />

<strong>Capnometer</strong>’s<br />

miniature size is proven in the<br />

most critical scenarios<br />

evaluate<br />

• Provides immediate information of respiratory status<br />

• Routine assessment of ventilation<br />

• Non-invasive assessment of respiratory status with mouthpiece<br />

• Validation and assessment of therapies and treatments<br />

• Use with mask or artificial airway<br />

resuscitate<br />

• Verification/Confirmation of tracheal intubation<br />

• Early indicator of return of spontaneous circulation<br />

(ROSC) during CPR<br />

• Informs the clinician of the effectiveness of cardiac<br />

compressions<br />

• Meets AHA 2005 Guidelines indicating that end-tidal CO 2<br />

monitoring is a safe and effective noninvasive indicator of<br />

cardiac output during CPR and may be an early indicator<br />

of ROSC in intubated patients<br />

Transport<br />

• Increases safety during patient transports by<br />

facilitating the integrity of an ET tube<br />

• Provides an indication of adequate ventilation<br />

• Helps monitor airway during patient movement<br />

ventilate<br />

• Assists in ongoing and routine assessment of ventilation<br />

(Helping to reduce the number of routine arterial blood gases)<br />

• Monitors the adequacy of ventilation during ventilator weaning<br />

• Early indicator of respiratory muscle failure<br />

<strong>Smiths</strong> <strong>Medical</strong> also provides a full-featured solution for emergency<br />

and transport ventilation resuscitation<br />

Pneupac ® VR1 Emergency and Transport Ventilator<br />

• Small, lightweight design has all the advantages of<br />

portability without compromising key clinical features<br />

• Auto/manual control allows the operator to choose<br />

automatic mode or to provide ventilation manually<br />

• Integrated patient demand system allows synchronized<br />

interaction with the patient’s own breathing efforts<br />

• Available with and without airmix


Capnocheck ®<br />

<strong>Capnometer</strong><br />

Technical Specifications<br />

• Description<br />

Pocket size, battery powered, quantitative<br />

capnometer for mainstream CO2 monitoring<br />

of adults and pediatrics<br />

• Measurements<br />

Non-dispersive IR absorption<br />

• Models<br />

Capnocheck ® WW8110-000<br />

Capnocheck ® with alarms WW8110-001<br />

• Warm Up<br />

Operational to full specification within 5 sec.<br />

• Calibration<br />

No routine calibration required<br />

• Certifications<br />

93/42/EEC and UL/CSA 60601-1<br />

• Dimensions<br />

2.1 x 1.5 x 1.5 inches (52 x 39 x 39 mm)<br />

• Weight<br />

2.1 oz (60 g) with batteries<br />

• Shock Absorption<br />

Withstands repeated 1 m drops<br />

PoWer reqUireMenTS<br />

• Batteries<br />

Two (2) AAA cell alkaline batteries (2 x 1.5VDC)<br />

(IEC Type LR03); 8 hour normal operation<br />

• Power Status<br />

LED Indicator<br />

environMenTAl<br />

• Operating Conditions<br />

Temperature: 23°F to 122°F (-5°C to +50°C)<br />

Humidity: 10 - 95% RH (non-condensing)<br />

Atmospheric Pressure: 70 - 120 kPa<br />

• Storage Conditions<br />

Temperature: -22°F to +158°F<br />

(-30°C to +70°C)<br />

Humidity: 5 - 100% RH (condensing)<br />

Atmospheric Pressure: 50 - 120 kPa<br />

DiSPlAYS<br />

• ETCO 2<br />

LED Numeric Display<br />

• Respiratory Rate<br />

LED Numeric Display<br />

• CO2 Bar Graph<br />

14 segment LED bar graph<br />

ConTrolS<br />

• Power<br />

ON key<br />

• Alarm Silence<br />

2 min. alarm silence key (WW8110-001 only)<br />

• ETCO 2<br />

Up/Down key for setting alarm limits<br />

(WW8110-001 only)<br />

Capnocheck ® <strong>Capnometer</strong> and Accessories<br />

Capnocheck ® <strong>Capnometer</strong> with Alarms (mmHg)......... WW8110-001SYS<br />

Capnocheck ® <strong>Capnometer</strong> (mmHg)............................... WW8110-000SYS<br />

Airway Adapter, box of 25................................................ WW8111<br />

Lanyard.............................................................................. WW8121<br />

Carrying Case................................................................... WW8120<br />

Airway Adapter Lanyard<br />

Carrying Case<br />

<strong>Smiths</strong> <strong>Medical</strong> PM, inc.<br />

Patient Monitoring and Ventilation<br />

N7W22025 Johnson Drive, Waukesha, Wisconsin 53186 USA<br />

Phone: 262-542-3100 Fax: 262-542-0718 Toll-Free: 800-558-2345<br />

www.smiths-medical.com/capnocheck<br />

<strong>Smiths</strong> <strong>Medical</strong>, part of the global technology business <strong>Smiths</strong> Group<br />

Product(s) described may not be licensed or available for sale outside of the United States.<br />

BCI, Capnocheck, Pneupac and the <strong>Smiths</strong> <strong>Medical</strong> design mark are trademarks of the <strong>Smiths</strong> <strong>Medical</strong> family of companies.<br />

The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other countries.<br />

All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners.<br />

© 2009 <strong>Smiths</strong> <strong>Medical</strong> family of companies. All rights reserved. BC2036 Rev.02 08/09<br />

Co2<br />

• Range<br />

0 - 99 mmHg (1)<br />

• Accuracy<br />

0-40 mmHg ± 2 mmHg, 41-99 mmHg 6%<br />

(2) (3)<br />

of reading, during standard conditions<br />

• Rise Time<br />

< 60 ms<br />

• Total System Response Time<br />

< 0.5 sec.<br />

noTeS<br />

(1) Gas reading showing actual partial pressure at<br />

current humidity level.<br />

(2) To include quantitative effect on gas reading<br />

from variations in environment conditions<br />

and presence of N2O, anesthetic agents,<br />

Ethanol, Isopropyl alcohol, He, Acetone and<br />

Methane, then the CO2 accuracy range should<br />

be increased to ±4 mmHg or 10% of reading,<br />

whichever is the greater.<br />

(3) CO2 was tested at a RR of 40. As RR rates<br />

increase above 60, the accuracy range will<br />

also increase.<br />

Information subject to change without notice. Please read the Instructions For Use supplied with the product for detailed instructions, warnings and cautions.<br />

For more information, please call <strong>Smiths</strong> <strong>Medical</strong> PM, Inc. at 262-542-3100 or 800-558-2345 or your <strong>Smiths</strong> <strong>Medical</strong> distributor.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!